pubmed:abstractText
The safety aspects of radiological methods continue to evolve. In this paper we report on two cases of skin burns in MRI caused by induced electrical current. A second- and a third-degree skin burn occurred during imaging in a 1.5 T system. The electromagnetic radiofrequency field inadvertently led to electrical currents caused by a conducting loop through the extremities and trunk. Skin burns induced by electrical current may occur in extremely rare cases even with standard MR imaging protocols operating within all current safety guidelines by inadvertently forming a closed conducting loop. By avoiding focal skin to skin contact of the extremities, this extremely rare adverse event can be avoided.
pubmed:language
ger
pubmed:journal
pubmed:citationSubset
IM
pubmed:status
MEDLINE
pubmed:month
Sep
pubmed:issn
0033-832X
pubmed:author
pubmed:issnType
Print
pubmed:volume
38
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
759-63
pubmed:dateRevised
2006-11-15
pubmed:meshHeading
pubmed:year
1998
pubmed:articleTitle
[Safety considerations to avoid current-induced skin burns in MRI procedures].
pubmed:affiliation
Forschungsschwerpunkt Radiologische Diagnostik und Therapie, Deutsches Krebsforschungszentrum (dkfz), Heidelberg.
pubmed:publicationType
Journal Article, English Abstract, Case Reports