1998-9-9
pubmed:abstractText
There are few established engineering guidelines aimed at reducing hemolysis for the design of centrifugal blood pumps. In this study, a fluid dynamic approach was applied to investigate hemolysis in centrifugal pumps. Three different strategies were integrated to examine the relationship between hemolysis and flow patterns. Hemolytic performances were evaluated in in vitro tests and compared with the flow patterns analyzed by flow visualization and computational fluid dynamic (CFD). Then our group tried to establish engineering guidelines to reduce hemolysis in the development of centrifugal blood pumps. The commercially available Nikkiso centrifugal blood pump (HPM-15) was used as a standard, and the dimensions of 2 types of gaps between the impeller and the casing, the axial and the radial gap, were varied. Four impellers with different vane outlet angles were also prepared and tested. Representative results of the hemolysis tests were as follows: The axial gaps of 0.5, 1.0, and 1.5 mm resulted in normalized index of hemolysis (NIH) values of 0.0028, 0.0013 and 0.0008 g/100 L, respectively. The radial gaps of 0.5 and 1.5 mm resulted in NIH values of 0.0012 and 0.0008 g/100 L, respectively. The backward type vane and the standard one resulted in NIH values of 0.0013 and 0.0002 g/100 L, respectively. These results revealed that small gaps led to more hemolysis and that the backward type vane caused more hemolysis. Therefore, the design parameters of centrifugal blood pumps could affect their hemolytic performances. In flow visualization tests, vortices around the impeller outer tip and tongue region were observed, and their patterns varied with the dimensions of the gaps. CFD analysis also predicted high shear stress consistent with the results of the hemolysis tests. Further investigation of the regional flow patterns is needed to discuss the cause of the hemolysis in centrifugal blood pumps.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:status
MEDLINE
pubmed:month
May
pubmed:issn
0160-564X
pubmed:author
pubmed:issnType
Print
pubmed:volume
22
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
393-8
pubmed:dateRevised
2007-11-15
pubmed:meshHeading
pubmed:year
1998
pubmed:articleTitle
Development of design methods of a centrifugal blood pump with in vitro tests, flow visualization, and computational fluid dynamics: results in hemolysis tests.
pubmed:affiliation
National Cardiovascular Center, Suita, Osaka, Japan.
pubmed:publicationType
Journal Article, Comparative Study, In Vitro, Research Support, Non-U.S. Gov't