Statements in which the resource exists as a subject.
PredicateObject
rdf:type
lifeskim:mentions
pubmed:issue
2
pubmed:dateCreated
1998-5-15
pubmed:abstractText
A fluorescence polarization immunoassay for barbiturates on the Abbott AxSYM analyzer is described. The assay displayed dilution linearity up to 1200 ng/mL; coefficients of variation varied from 5.96 to 8.61%; recovery varied from 94.9 to 105.3%; and sensitivity was less than 40 ng/mL. Good correlation between the standard six- and factory two-point calibration methods was observed. The Immunoassay demonstrated good cross-reactivity to several commonly prescribed barbiturates; low cross-reactivity with structurally similar compounds; low interference from endogenous substances, dyes, preservatives, and several commonly available adulterants; and good correlation with the TDx Barbiturate Urine assay.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:chemical
pubmed:status
MEDLINE
pubmed:issn
0146-4760
pubmed:author
pubmed:issnType
Print
pubmed:volume
22
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
105-11
pubmed:dateRevised
2006-11-15
pubmed:meshHeading
pubmed:articleTitle
A barbiturate screening assay for the Abbott AxSYM analyzer.
pubmed:affiliation
Department of Chemistry, Abbott Laboratories, Abbott Park, Illinois 60064-3500, USA.
pubmed:publicationType
Journal Article, Comparative Study