Statements in which the resource exists as a subject.
PredicateObject
rdf:type
lifeskim:mentions
pubmed:issue
1
pubmed:dateCreated
1998-3-5
pubmed:abstractText
This report documents clinical and hemodynamic benefits of balloon mitral valvuloplasty (BMV) using a bifoil balloon catheter from a single center in 415 consecutive cases of rheumatic mitral stenosis (MS). The procedure was successful in 396 (95.2%) patients, with an increase in mitral valve area (MVA) from 0.82+/-0.35 cm2 to 2.21+/-0.24 cm2 (P < 0.001). There were 2 (0.48%) in-hospital deaths, and 6 (1.44%) patients developed acute mitral regurgitation. The procedural and fluroscopy time was reduced significantly from 52+/-11 and 38+/-6 min to 33+/-7 and 19+/-5 min, respectively, after modifications of technique in our last 326 cases. The bifoil balloon catheter technique is safe and effective, and provides excellent hemodynamic benefits which are sustained at long-term follow-up. This technique should be considered as an addition to the existing armamentarium of interventional cardiologists performing mitral balloon valvuloplasty.
pubmed:commentsCorrections
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:status
MEDLINE
pubmed:month
Jan
pubmed:issn
0098-6569
pubmed:author
pubmed:issnType
Print
pubmed:volume
43
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
43-7
pubmed:dateRevised
2010-11-18
pubmed:meshHeading
pubmed:year
1998
pubmed:articleTitle
Balloon mitral valvuloplasty with bifoil catheter: immediate and long-term follow-up results.
pubmed:affiliation
Apollo Heart Institute, Apollo Hospital, Hyderabad, Andhra Pradesh, India.
pubmed:publicationType