As stated in a recent document of the National Committee for Clinical Laboratory Standards (NCCLS Document C42-P, April 1995), expression of the results of erythrocyte zinc protoporphyrin (ZPP) in different units is a main source of confusion which may hamper the wide use of this test. The possible inter-conversion of the ZPP values expressed as microgram/g Hb, microgram/dL blood and microgram/dL red cells was assessed. It was found that in all cases, despite a good correlation between these values (r > or = 0.939), the error of the estimation was higher than allowed by the American College of Pathologists criteria. Such findings show that ZPP values expressed in different units cannot be inter-converted as a function of the corresponding regression equation. Similarly, the use of hematofluorometers with an analogic-digital converter to express results as microgram ZPP/ dL blood might pose an additional source of error.
|
pubmed:language |
spa
|
pubmed:journal | |
pubmed:citationSubset |
IM
|
pubmed:chemical | |
pubmed:status |
MEDLINE
|
pubmed:month |
Oct
|
pubmed:issn |
0036-4355
|
pubmed:author | |
pubmed:issnType |
Print
|
pubmed:volume |
42
|
pubmed:owner |
NLM
|
pubmed:authorsComplete |
Y
|
pubmed:pagination |
415-8
|
pubmed:dateRevised |
2006-11-15
|
pubmed:meshHeading | |
pubmed:year |
1997
|
pubmed:articleTitle |
[Interconversion of the results for erythrocyte zinc protoporphyrin expressed in different units].
|
pubmed:affiliation |
Laboratorio Central, Hospital General de Galicia-Clínico Universitario, Santiago de Compostela.
|
pubmed:publicationType |
Journal Article,
English Abstract
|