Statements in which the resource exists as a subject.
PredicateObject
rdf:type
lifeskim:mentions
pubmed:issue
6 Suppl
pubmed:dateCreated
1998-3-17
pubmed:abstractText
The objective of this study was to determine whether insertion length and number of active channels remained the same after reimplantation of a cochlear implant. A retrospective case review of 170 consecutively implanted multichannedl cochlear implants was conducted. Seventeen of these devices had to be replaced. Data were analyzed for the Nucleus cochlear implant users who were reimplanted in the same ear. For most subjects, insertion length and number of channels remained unchanged, but a few subjects experienced substantial decreases. When the whole group was considered, a small but statistically significant drop was noted for both parameters. In conclusion, although reimplantation is technically possible, the first procedure provides the optimal surgical environment.
pubmed:grant
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:status
MEDLINE
pubmed:month
Nov
pubmed:issn
0192-9763
pubmed:author
pubmed:issnType
Print
pubmed:volume
18
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
S60-1
pubmed:dateRevised
2007-11-14
pubmed:meshHeading
pubmed:year
1997
pubmed:articleTitle
Cochlear implant reimplantation.
pubmed:affiliation
Department of Otolaryngology-Head and Neck Surgery, Indiana University School of Medicine, Indianapolis 46202, USA.
pubmed:publicationType
Journal Article, Research Support, U.S. Gov't, P.H.S.