pubmed:abstractText
Direct and alternating current electrochemical tests were carried out on Ti6Al4V with different surface finishing and with hydroxyapatite (HA) coatings. Sand-blasting and rough titanium deposits obtained by vacuum plasma spraying (VPS) bring about an increase of passive and corrosion current density (c.d.) with respect to smooth Ti6Al4V, as a consequence of the augmentation of the real surface. The presence of HA deposits obtained by VPS causes an increase of passive and corrosion c.d. of the metallic substrate of about one order of magnitude and this should be taken into account in view of human body applications.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:chemical
pubmed:status
MEDLINE
pubmed:month
Jun
pubmed:issn
0142-9612
pubmed:author
pubmed:issnType
Print
pubmed:volume
18
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
783-7
pubmed:dateRevised
2004-11-17
pubmed:meshHeading
pubmed:year
1997
pubmed:articleTitle
Effect of different surface finishing and of hydroxyapatite coatings on passive and corrosion current of Ti6Al4V alloy in simulated physiological solution.
pubmed:affiliation
Dipartimento di Chimica Fisica Applicata, Politecnico di Milano, Italy.