The construction, calibration, and surgical placement of a device for monitoring uterine motility are described. The device, a linear stretch gauge, consisted of a length of flexible tubing filled with mercury and connected at both ends to copper wire leads. An increase in the length of the mercury-filled tubing caused a change in resistance. This change was quantitated, using a modified Wheatstone bridge circuit. In in vitro test, the stretch gauges demonstrated linear response in millivolt output to elongation over a range that was comparable to physiologic responses observed with the gauges placed in vivo. If surgically placed around 1 uterine horn, stretch gauges responded to uterine contractile events with specificity and sensitivity. Calibration of the device at time of placement permitted monitoring for possible increases in uterine circumference.
|
pubmed:language |
eng
|
pubmed:journal | |
pubmed:citationSubset |
IM
|
pubmed:chemical | |
pubmed:status |
MEDLINE
|
pubmed:month |
Aug
|
pubmed:issn |
0002-9645
|
pubmed:author | |
pubmed:issnType |
Print
|
pubmed:volume |
38
|
pubmed:owner |
NLM
|
pubmed:authorsComplete |
Y
|
pubmed:pagination |
1275-8
|
pubmed:dateRevised |
2003-11-14
|
pubmed:meshHeading | |
pubmed:year |
1977
|
pubmed:articleTitle |
A technique for chronic, extraluminal measurement of uterine activity.
|