pubmed:abstractText |
The authors report a case of massive wear of carbon fiber-reinforced polyethylene used as an insert in a knee prosthesis. The finding, which was studied under a scanning electron microscope, in agreement with mechanical resistance and laboratory testing reported in the literature, confirms that this composite material (poly-two) is not more advantageous to use that ultra-high molecular weight polyethylene in terms of wear, despite favorable experimental premises.
|
pubmed:language |
eng
|
pubmed:journal | |
pubmed:citationSubset |
IM
|
pubmed:chemical | |
pubmed:status |
MEDLINE
|
pubmed:issn |
0009-4749
|
pubmed:author | |
pubmed:issnType |
Print
|
pubmed:volume |
81
|
pubmed:owner |
NLM
|
pubmed:authorsComplete |
Y
|
pubmed:pagination |
263-7
|
pubmed:dateRevised |
2010-6-18
|
pubmed:meshHeading |
pubmed-meshheading:9009408-Aged,
pubmed-meshheading:9009408-Carbon,
pubmed-meshheading:9009408-Female,
pubmed-meshheading:9009408-Follow-Up Studies,
pubmed-meshheading:9009408-Humans,
pubmed-meshheading:9009408-Knee Prosthesis,
pubmed-meshheading:9009408-Polyethylenes,
pubmed-meshheading:9009408-Prostheses and Implants,
pubmed-meshheading:9009408-Prosthesis Failure,
pubmed-meshheading:9009408-Time Factors
|
pubmed:articleTitle |
Wear in carbon fiber-reinforced polyethylene (poly-two) knee prostheses.
|
pubmed:affiliation |
Clinica Ortopedica dell'Università, Istituti Ortopedici Rizzoli, Bologna.
|
pubmed:publicationType |
Journal Article,
Case Reports
|