pubmed:abstractText
The efficacy of an innovative intravenous membrane oxygenator (IMO) was tested acutely (6-8 hrs) in seven calves. The IMO prototypes consisted of a central polyurethane balloon within a bundle of hollow fibers with a membrane surface area of 0.14 m2. The IMO devices were inserted through the external jugular vein into the inferior vena cava of anesthetized calves (68.9 +/- 2.3 kg). Rhythmic balloon pulsation (60-120 bpm) was controlled with an intra-aortic balloon pump console. Oxygen sweep gas was delivered through the device at 3.0 L/min. Gas concentrations were monitored continuously by mass spectroscopy. The principal results were as follows: 1) oxygen and carbon dioxide exchange ranged from 125 to 150 ml/min/m2 and 150 to 200 ml/min/m2, respectively; 2) there was at least a 30-50% augmentation of gas exchange with balloon pulsation; 3) maximum exchange occurred with 60-90 bpm balloon pulsations; and 4) hemodynamic parameters remained unchanged. There were no device related complications, and the feasibility of insertion of the device by a cervical cut-down was established. These acute in vivo experiments show that the Pittsburgh IMO device can exchange oxygen and carbon dioxide gases in vivo at levels consistent with this current prototype design, and that intravenous balloon pulsation significantly enhances gas exchange without causing any end-organ damage.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:chemical
pubmed:status
MEDLINE
pubmed:issn
1058-2916
pubmed:author
pubmed:issnType
Print
pubmed:volume
42
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
M609-15
pubmed:dateRevised
2007-11-15
pubmed:meshHeading
pubmed:articleTitle
Acute in vivo studies of the Pittsburgh intravenous membrane oxygenator.
pubmed:affiliation
Department of Surgery, University of Pittsburgh, Pennsylvania 15219, USA.
pubmed:publicationType