pubmed:abstractText
The radiation dose to physicians' hands without and with use of needle holders was determined at 10 computed tomography (CT) fluoroscopy-guided transthoracic needle biopsies. As measured with ionization chambers (tube voltage, 80 kVp; tube current, 30 mA), the mean absorbed dose rate without and with holders was 1.14 mGy/sec +/- 0.02 (standard deviation) and 0.019 mGy/sec +/- 0.001, respectively. The mean duration of CT fluoroscopy was 59 seconds in 10 biopsies performed with a holder and 82 seconds in 10 biopsies performed without a holder (difference not statistically significant). The needle holders did not cause any artifacts that interfered with the biopsy procedure.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
AIM
pubmed:status
MEDLINE
pubmed:month
Nov
pubmed:issn
0033-8419
pubmed:author
pubmed:issnType
Print
pubmed:volume
201
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
576-8
pubmed:dateRevised
2004-11-17
pubmed:meshHeading
pubmed:year
1996
pubmed:articleTitle
Radiation dosimetry at CT fluoroscopy: physician's hand dose and development of needle holders.
pubmed:affiliation
Department of Radiology, Fujita Health University School of Medicine, Aichi, Japan.
pubmed:publicationType
Journal Article