pubmed:abstractText |
This study analyzes the signal-to-noise ratio (SNR) in magnetic resonance imaging. The factors that determine the SNR are derived starting from basic principles. The SNR, for a given object, is shown to be proportional to the voxel volume and the square root of the acquisition time. The noise generated by the body is derived using a cylindrical model and is shown to be proportional to the square of the radius and the square root of the length.
|
pubmed:language |
eng
|
pubmed:journal | |
pubmed:citationSubset |
IM
|
pubmed:status |
MEDLINE
|
pubmed:month |
Sep
|
pubmed:issn |
0740-3194
|
pubmed:author | |
pubmed:issnType |
Print
|
pubmed:volume |
36
|
pubmed:owner |
NLM
|
pubmed:authorsComplete |
Y
|
pubmed:pagination |
494-7
|
pubmed:dateRevised |
2004-11-17
|
pubmed:meshHeading | |
pubmed:year |
1996
|
pubmed:articleTitle |
Noise in MRI.
|