pubmed:abstractText |
We have developed an autoflow control system for the Terumo Capiox centrifugal pump, the SP-101. First, we made an external control system regulated by a personal computer through a serial communications line. Although an in vivo experiment using a goat showed excellent performance in regulating the flow, a communication error that caused an abrupt and instantaneous flow fluctuation was noted occasionally. Therefore, we incorporated a check-sum function to solve this problem. After this improvement, 5 patients safely underwent cardiopulmonary bypass using a pump having this autoflow control. As the next step for more practical usage, we developed a built-in autoflow control system in which the control device is entirely incorporated into the console without a communications line. This system demonstrated better response and reliability than the external control system and was applied successfully in 3 patients. In conclusion, this autoflow mode will expand the usage of centrifugal pumps through improved handling.
|
pubmed:language |
eng
|
pubmed:journal | |
pubmed:citationSubset |
IM
|
pubmed:status |
MEDLINE
|
pubmed:month |
Jun
|
pubmed:issn |
0160-564X
|
pubmed:author | |
pubmed:issnType |
Print
|
pubmed:volume |
20
|
pubmed:owner |
NLM
|
pubmed:authorsComplete |
Y
|
pubmed:pagination |
625-31
|
pubmed:dateRevised |
2006-11-15
|
pubmed:meshHeading |
pubmed-meshheading:8817968-Algorithms,
pubmed-meshheading:8817968-Cardiopulmonary Bypass,
pubmed-meshheading:8817968-Centrifugation,
pubmed-meshheading:8817968-Computer Simulation,
pubmed-meshheading:8817968-Equipment Design,
pubmed-meshheading:8817968-Equipment Safety,
pubmed-meshheading:8817968-Heart-Assist Devices,
pubmed-meshheading:8817968-Humans,
pubmed-meshheading:8817968-Quality Control,
pubmed-meshheading:8817968-Rheology
|
pubmed:year |
1996
|
pubmed:articleTitle |
Clinical application of a newly developed autoflow control system for the Terumo centrifugal pump: from external control to built-in direct control.
|
pubmed:affiliation |
Department of Cardiovascular Surgery, Heart Institute of Japan, Tokyo Women's Medical College, Japan.
|
pubmed:publicationType |
Journal Article,
Research Support, Non-U.S. Gov't
|