Statements in which the resource exists as a subject.
PredicateObject
rdf:type
lifeskim:mentions
pubmed:issue
4
pubmed:dateCreated
1997-1-14
pubmed:abstractText
We demonstrate an interleaved dual spin echo-based sequence for quantitative measurement of Magnetisation Transfer Ratio (MTR) in a clinical environment that overcomes the problems of patient motion between scans faced by noninterleaved methods. The sequence also provides proton density and T2-weighted images, allowing direct comparison among the three contrast regimes. Phantom studies and in vivo measurements on normal controls show the sequence to be robust in normal use. The values of MTR calculated from the sequence are shown to be precise and reproducible enough to allow regional variations to be identified within and between white matter and other brain tissues.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:status
MEDLINE
pubmed:issn
0730-725X
pubmed:author
pubmed:issnType
Print
pubmed:volume
14
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
403-11
pubmed:dateRevised
2006-11-15
pubmed:meshHeading
pubmed:year
1996
pubmed:articleTitle
An interleaved sequence for accurate and reproducible clinical measurement of magnetization transfer ratio.
pubmed:affiliation
Institute of Neurology, London, UK.
pubmed:publicationType
Journal Article, Research Support, Non-U.S. Gov't