pubmed:abstractText
The transtrochanteric surgical approach to the hip is commonly used, especially for revision hip surgery. Failure of the trochanter to heal can lead to hardware failure, persistent pain, and limp. Rigid internal fixation is needed in this approach to achieve an adequate rate of healing. Newer cable and cable grip systems have been designed to improve trochanteric fixation, but have not been compared to the older Charnley wire fixation techniques. In this study, an in vitro mechanical method previously used to test wire fixation methods was used to compare wire, cable, and cable grip fixation methods. A quasistatic mechanical distraction device was used to compare structural stiffness, load to clinical failure, energy to clinical failure, and maximum load resisted by the fixation devices. The cable grip system was found to be stiffer, to resist a larger load to mechanical failure at 1-cm displacement, and to absorb a greater amount of energy to clinical failure when compared with the other systems. These data suggest that use of the cable grip fixation method should result in improved clinical success compared with the Charnley wire technique.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
AIM
pubmed:status
MEDLINE
pubmed:month
Aug
pubmed:issn
0009-921X
pubmed:author
pubmed:issnType
Print
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
317-25
pubmed:dateRevised
2006-11-15
pubmed:meshHeading
pubmed:year
1996
pubmed:articleTitle
Comparison of the mechanical performance of trochanteric fixation devices.
pubmed:affiliation
Department of Orthopaedics, The University of Texas Health Science Center at San Antonio, USA.
pubmed:publicationType
Journal Article, Comparative Study