pubmed:abstractText
A gas chromatographic method using nitrogen-phosphorus detection was developed to quantitate clozapine in plasma or serum. Methyl clonazepam was used as an internal standard. Sample preparation included a single-step extraction with ethyl acetate, which was injected directly onto a wide-bore capillary column. Within-run and total precision, measured as percent coefficient of variation, were determined at low, therapeutic, and high clozapine plasma concentrations. The within-run precision for the low, therapeutic, and high clozapine plasma samples was 5.2, 2.7, and 2.4%, respectively. The total precision for the low, therapeutic, and high clozapine plasma samples was 10.0, 2.6, and 2.0%, respectively. Analytical accuracy was evaluated by comparing quantitative results with those obtained from a reference laboratory. Those samples containing therapeutic or high concentrations agreed within 3%; the sample containing a subtherapeutic concentration differed by 11.9%. The limit of quantitation was determined to be 35 ng/mL, and the upper limit of linearity was 3000 ng/mL. No significant interferences were detected after testing more than two dozen drugs and metabolites.
pubmed:commentsCorrections
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:chemical
pubmed:status
MEDLINE
pubmed:issn
0146-4760
pubmed:author
pubmed:issnType
Print
pubmed:volume
19
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
537-41
pubmed:dateRevised
2004-11-17
pubmed:meshHeading
pubmed:articleTitle
A rapid gas chromatographic method quantitating clozapine in human plasma or serum for the purpose of therapeutic monitoring.
pubmed:affiliation
Associated Regional and University Pathologists, Inc., Salt Lake City, UT 84108, USA.
pubmed:publicationType
Journal Article