In Germany, the target values for External Quality Assessment (EQA) and internal accuracy control are determined by Reference Methods for several analytes, including cholesterol, creatinine, uric acid, and glucose. We present candidate Reference Methods for these compounds, based on isotope dilution-gas chromatography--mass spectrometry methods that have been developed at INSTAND, one of the two official Germany EQA reference institutions. Each Reference Method target value is calculated from six independent measurements performed on three different days. The mean method CVs ranged from 0.66% for glucose to 0.96% for creatinine. The inaccuracy (bias) of the methods is < 0.7%, as compared with the Standard Reference Material 909 of the National Institute of Standards and Technology. The maximum total error of a Reference Method value, including the 95% confidence interval and systematic errors, is < 2.3%. The presented candidate Reference Methods have been successfully used to set target values in the German EQA scheme and the internal accuracy control of routine laboratories.
pubmed:commentsCorrections
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:chemical
pubmed:status
MEDLINE
pubmed:month
Jun
pubmed:issn
0009-9147
pubmed:author
pubmed:issnType
Print
pubmed:volume
39
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
993-1000
pubmed:dateRevised
2006-11-15
pubmed:meshHeading
pubmed:year
1993
pubmed:articleTitle
Candidate reference methods for determining target values for cholesterol, creatinine, uric acid, and glucose in external quality assessment and internal accuracy control. I. Method setup.
pubmed:affiliation
Institut für Standardisierung und Dokumentation im medizinischen Laboratorium e.V. INSTAND e.V., Düsseldorf, Germany.
pubmed:publicationType
Journal Article