pubmed:abstractText
A self-tuning, nonlinear controller was developed to drive portable or implantable micro-pumps for blood glucose control in diabetic subjects. The parameter estimation is based on a recursive least-square algorithm applied to a discrete time simplified mathematical model of the glucose system; the controller, which is integrated with the estimator, uses an extended minimum-variance method. The software was designed to drive a micro-controller wearable unit, operating with different kinds of glucose sensors, in order to perform chemical and biological experimentations. Some strategies were introduced to avoid hypoglycaemia, even tolerating a reduction in control speed and accuracy. The whole system was tested in a simulation study, performed on a mathematical model implemented on a personal computer. The tests were performed while simulating different controller structures and settings, and patient responses. They showed a satisfactory control behaviour, mostly as far as stability and robustness are concerned, in all simulated conditions.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:status
MEDLINE
pubmed:month
Jan
pubmed:issn
0391-3988
pubmed:author
pubmed:issnType
Print
pubmed:volume
16
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
51-7
pubmed:dateRevised
2006-11-15
pubmed:meshHeading
pubmed:year
1993
pubmed:articleTitle
A simulation study on a self-tuning portable controller of blood glucose.
pubmed:affiliation
Medical Pathology Institute, Bioengineering Laboratory, University of Perugia, Italy.