rdf:type |
|
lifeskim:mentions |
|
pubmed:issue |
1
|
pubmed:dateCreated |
1993-4-19
|
pubmed:abstractText |
An inexpensive and easy method of calibration of brachytherapy 137Cs tubes is described. Cesium brachytherapy sources are calibrated relative to an NIST-calibrated source. The detection system is a NaI detector and single channel analyzer with the window set on the photo peak.
|
pubmed:language |
eng
|
pubmed:journal |
|
pubmed:citationSubset |
IM
|
pubmed:chemical |
|
pubmed:status |
MEDLINE
|
pubmed:issn |
0094-2405
|
pubmed:author |
|
pubmed:issnType |
Print
|
pubmed:volume |
20
|
pubmed:owner |
NLM
|
pubmed:authorsComplete |
Y
|
pubmed:pagination |
245-6
|
pubmed:dateRevised |
2008-11-21
|
pubmed:meshHeading |
|
pubmed:articleTitle |
Routine verification of strength of 137Cs brachytherapy sources using a NaI detector.
|
pubmed:affiliation |
Department of Radiation Medicine, University of Kentucky Medical Center, Lexington 40536-0084.
|
pubmed:publicationType |
Journal Article
|