Statements in which the resource exists as a subject.
PredicateObject
rdf:type
lifeskim:mentions
pubmed:issue
1
pubmed:dateCreated
1993-4-19
pubmed:abstractText
An inexpensive and easy method of calibration of brachytherapy 137Cs tubes is described. Cesium brachytherapy sources are calibrated relative to an NIST-calibrated source. The detection system is a NaI detector and single channel analyzer with the window set on the photo peak.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:chemical
pubmed:status
MEDLINE
pubmed:issn
0094-2405
pubmed:author
pubmed:issnType
Print
pubmed:volume
20
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
245-6
pubmed:dateRevised
2008-11-21
pubmed:meshHeading
pubmed:articleTitle
Routine verification of strength of 137Cs brachytherapy sources using a NaI detector.
pubmed:affiliation
Department of Radiation Medicine, University of Kentucky Medical Center, Lexington 40536-0084.
pubmed:publicationType
Journal Article