pubmed:abstractText
During implantation of epicardial automatic defibrillator systems, occasional patients have difficulty in obtaining adequate defibrillation thresholds. Of 236 consecutive patients undergoing implantation of epicardial defibrillator systems, 18 patients received a 3-patch (n = 15) or 4-patch (n = 3) defibrillator system. Twelve patients who received a multiple-patch defibrillator system had a best 2-patch defibrillation energy requirement of > or = 30 J; in the remaining 6 patients less stringent clinical criteria were used in the decision to add a third defibrillator patch (defibrillation energy requirement > 18 J in 4 patients, and > 20 J in 2 patients). Technically, multiple-patch systems were made possible with either the use of Y-connectors or defibrillators allowing output to 3 patches. In 3 patients, addition of a third epicardial patch still resulted in a defibrillation energy requirement of > or = 30 J; in these 3 patients, addition of a fourth patch resulted in a defibrillation energy requirement of < or = 20 J. All patients receiving a multiple-patch defibrillator system had a reduction in defibrillation energy requirement, and 12 patients had a reduction in defibrillation energy requirement of > or = 10 J over the best 2-patch defibrillation energy requirement. In the patients who eventually had placement of a multiple-patch system, the best 2-patch defibrillation energy requirement was > 18 J in 4 patients, > 20 J in 2 patients, > or = 30 J in 9 patients, and > 40 J in 3 patients.(ABSTRACT TRUNCATED AT 250 WORDS)
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
AIM
pubmed:chemical
pubmed:status
MEDLINE
pubmed:month
Jan
pubmed:issn
0002-9149
pubmed:author
pubmed:issnType
Print
pubmed:day
1
pubmed:volume
71
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
68-71
pubmed:dateRevised
2005-11-17
pubmed:meshHeading
pubmed:year
1993
pubmed:articleTitle
Use of multiple patches during implantation of epicardial defibrillator systems.
pubmed:affiliation
Department of Medicine, Loyola University Medical Center, Maywood, Illinois 60153.
pubmed:publicationType
Journal Article