Statements in which the resource exists as a subject.
PredicateObject
rdf:type
lifeskim:mentions
pubmed:issue
2
pubmed:dateCreated
1993-9-30
pubmed:abstractText
The application of automated blood pressure measurement during exercise has been limited by inaccuracies introduced by the effects of accompanying motion and noise. We evaluated a newly developed automated blood pressure monitor for measuring exercise blood pressure (Colin STBP-680; Colin, San Antonio, Texas, USA). The STBP-680 uses acoustic transduction with the assistance of the electrocardiogram R-wave to trigger the sampling period for blood pressure measurement. The automated monitor readings were compared with simultaneous technician mercury sphygmomanometric readings in the same arm. Blood pressure was measured in 18 men at rest and during exercise at 40% VO2 peak, (low intensity), 70% VO2 peak (moderate intensity) and VO2 peak (high intensity) on the cycle ergometer. Mean(s.d.) systolic blood pressure difference between the automated monitor and mercury manometer readings at rest and during exercise at low, moderate and high work intensities were 3(0) mmHg, 3(2) mmHg, 1(1) mmHg, and 0(11) mmHg respectively (analysis of variance; P > 0.05). Resting diastolic blood pressure obtained with the STBP-680 was similar to the mercury manometer readings (78(10) versus 81(7) mmHg (P > 0.05). Exercise diastolic pressure at the low level of work intensity was almost identical between the automated monitor and mercury manometer readings (64(8) versus 65(10) mmHg (not significant)). Diastolic blood pressure readings between the STBP-680 and mercury manometer showed a greater difference at the moderate and high workloads (11 mmHg and 9 mmHg, respectively), but this difference was not significant (P > 0.05).(ABSTRACT TRUNCATED AT 250 WORDS)
pubmed:commentsCorrections
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:status
MEDLINE
pubmed:month
Jun
pubmed:issn
0306-3674
pubmed:author
pubmed:issnType
Print
pubmed:volume
27
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
107-9
pubmed:dateRevised
2009-11-18
pubmed:meshHeading
pubmed:year
1993
pubmed:articleTitle
Evaluation of the Colin STBP-680 at rest and during exercise: an automated blood pressure monitor using R-wave gating.
pubmed:affiliation
Department of Human Performance and Sport Studies, University of Tennessee, Knoxville.
pubmed:publicationType
Journal Article