pubmed-article:8273877pubmed:abstractTextThe first 2000 incidents reported to the Australian INcident Monitoring Study were analysed with respect to the role of the oxygen analyser; 27 (1%) were first detected by the oxygen analyser. All of these were amongst the 1256 incidents which occurred in association with general anaesthesia, of which 48% were "human detected" and 52% "monitor detected". The oxygen analyser was ranked 7th and detected 4% of these monitor detected incidents. This figure would have been much higher had the oxygen analyser been correctly used on more occasions. The oxygen analyser detected 10 ventilator-driving-gas leaks into the circuit, 6 hypoxic mixtures due to rotameter settings, 3 inappropriate nitrous oxide concentrations, 2 disconnections and 1 leak at the common gas outlet, and 2 partial and 1 total failure of ventilation. In a theoretical analysis of these 1256 incidents it was considered that the oxygen analyser, used on its own, would have detected 114 (9%), had they been allowed to evolve (3% before any potential for organ damage). In 4 incidents an oxygen analyser gave faulty readings, in 3 caused a leak and in one a total circuit obstruction; 5 incidents were not detected because the alarm had been disabled. Despite the advent of piped gas supplies, failure of gas delivery or delivery of a "wrong" gas mixture still occurs surprisingly frequently in current anaesthetic practice; hypoxic mixtures were supplied on 16 occasions, other "wrong" mixtures on 23 and the oxygen supply failed on 7 occasions.(ABSTRACT TRUNCATED AT 250 WORDS)lld:pubmed
pubmed-article:8273877pubmed:languageenglld:pubmed
pubmed-article:8273877pubmed:journalhttp://linkedlifedata.com/r...lld:pubmed
pubmed-article:8273877pubmed:citationSubsetIMlld:pubmed
pubmed-article:8273877pubmed:chemicalhttp://linkedlifedata.com/r...lld:pubmed
pubmed-article:8273877pubmed:statusMEDLINElld:pubmed
pubmed-article:8273877pubmed:monthOctlld:pubmed
pubmed-article:8273877pubmed:issn0310-057Xlld:pubmed
pubmed-article:8273877pubmed:authorpubmed-author:VEISJJlld:pubmed
pubmed-article:8273877pubmed:authorpubmed-author:Van der...lld:pubmed
pubmed-article:8273877pubmed:authorpubmed-author:RuncimanW BWBlld:pubmed
pubmed-article:8273877pubmed:authorpubmed-author:BarkerLLlld:pubmed
pubmed-article:8273877pubmed:issnTypePrintlld:pubmed
pubmed-article:8273877pubmed:volume21lld:pubmed
pubmed-article:8273877pubmed:ownerNLMlld:pubmed
pubmed-article:8273877pubmed:authorsCompleteYlld:pubmed
pubmed-article:8273877pubmed:pagination570-4lld:pubmed
pubmed-article:8273877pubmed:dateRevised2006-8-28lld:pubmed
pubmed-article:8273877pubmed:meshHeadingpubmed-meshheading:8273877-...lld:pubmed
pubmed-article:8273877pubmed:meshHeadingpubmed-meshheading:8273877-...lld:pubmed
pubmed-article:8273877pubmed:meshHeadingpubmed-meshheading:8273877-...lld:pubmed
pubmed-article:8273877pubmed:meshHeadingpubmed-meshheading:8273877-...lld:pubmed
pubmed-article:8273877pubmed:meshHeadingpubmed-meshheading:8273877-...lld:pubmed
pubmed-article:8273877pubmed:meshHeadingpubmed-meshheading:8273877-...lld:pubmed
pubmed-article:8273877pubmed:meshHeadingpubmed-meshheading:8273877-...lld:pubmed
pubmed-article:8273877pubmed:meshHeadingpubmed-meshheading:8273877-...lld:pubmed
pubmed-article:8273877pubmed:meshHeadingpubmed-meshheading:8273877-...lld:pubmed
pubmed-article:8273877pubmed:year1993lld:pubmed
pubmed-article:8273877pubmed:articleTitleThe Australian Incident Monitoring Study. The oxygen analyser: applications and limitations--an analysis of 200 incident reports.lld:pubmed
pubmed-article:8273877pubmed:affiliationDepartment of Anaesthesia and Intensive Care, University of Adelaide, S.A.lld:pubmed
pubmed-article:8273877pubmed:publicationTypeJournal Articlelld:pubmed