Statements in which the resource exists as a subject.
PredicateObject
rdf:type
lifeskim:mentions
pubmed:issue
9
pubmed:dateCreated
1993-12-21
pubmed:abstractText
Serial reticulocyte counts on a patient with an unstable haemoglobin haemolytic anaemia whose red cells contained Heinz bodies were obtained by two automated methods. One false low and several false high counts were obtained with standard threshold settings; each was accompanied by activation of the appropriate alarm. With the FACScan a bimodal red cell distribution was found, corresponding to the proportions of Heinz body positive and negative cells. This case illustrates the fact that Heinz bodies can interfere with automated reticulocyte counting methods, and to a degree that could be clinically important.
pubmed:commentsCorrections
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
AIM
pubmed:status
MEDLINE
pubmed:month
Sep
pubmed:issn
0021-9746
pubmed:author
pubmed:issnType
Print
pubmed:volume
46
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
878-9
pubmed:dateRevised
2009-11-18
pubmed:meshHeading
pubmed:year
1993
pubmed:articleTitle
Errors in automated reticulocyte counts due to Heinz bodies.
pubmed:affiliation
Section of Paediatric Haematology, University of Sheffield.
pubmed:publicationType