pubmed:abstractText
The University of Cincinnati has reproduced the original formulation for the Livermore Thoracic Phantom lungs using contemporary materials and has adopted the linear attenuation coefficient as the primary quality assurance parameter for evaluating the performance capabilities of these new lung phantoms. The Livermore Thoracic Phantom was originally fabricated in 1978 to intercalibrate detector systems used to measure plutonium and other low-energy, photon emitting radionuclides deposited in the respiratory tract. The linear attenuation coefficient is a critical performance indicator for these phantom lungs since the presence of any material with a high effective atomic number (where Z > or = 20) will make a significant change in the photoelectric cross section, the predominant mode of interaction for plutonium x rays. A set of test lungs was fabricated with KCl to introduce a known quantity of 40K in the phantom and to determine, by measurement and calculations, what change would be made to the attenuation coefficient at photon energies below 100 keV as a result of the modified formulation. The KCl increased the linear attenuation coefficient below 60 keV by more than a factor of two, which would produce a substantial systematic error in any subsequent calibration measurements performed with these modified phantom lungs. These results support use of the attenuation coefficient as an important performance indicator for the Livermore Thoracic Phantom lungs and also suggest that KCl not be added to the lung tissue substitute formulation as a means to incorporate 40K in the phantom for low energy calibrations.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:chemical
pubmed:status
MEDLINE
pubmed:month
Jul
pubmed:issn
0017-9078
pubmed:author
pubmed:issnType
Print
pubmed:volume
67
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
39-46
pubmed:dateRevised
2004-11-17
pubmed:meshHeading
pubmed:year
1994
pubmed:articleTitle
Measurement of the attenuation coefficient for Livermore Thoracic Phantom lungs fabricated using contemporary materials.
pubmed:affiliation
University of Cincinnati, OH 45221-0072.
pubmed:publicationType
Journal Article