pubmed:abstractText
We have described the design and design considerations of the desflurane Tec 6 "vaporizer" and have tested its performance characteristics. The vaporizer differs from previous vaporizers designed for anaesthesia in that electromechanical rather than mechanical controls accommodate the different physical characteristics of desflurane. This design, while offering perhaps an increased risk of failure (owing to sophisticated electronic components and circuitry), on the other hand offers the decreased likelihood of accidental delivery of very large concentrations of liquid anaesthetic resulting from tilting or overfilling and alarms and warnings not previously incorporated into the design of anaesthetic vaporizers. The output characteristics of the vaporizer are as expected, based on the design: desflurane concentration output in oxygen has accuracy (+/- 15%) which is similar to that of the mechanical vaporizers; output decreases when nitrous oxide is added owing to the lower viscosity, but remains within 20% of the dial setting or 0.5% absolute.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:chemical
pubmed:status
MEDLINE
pubmed:month
Apr
pubmed:issn
0007-0912
pubmed:author
pubmed:issnType
Print
pubmed:volume
72
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
474-9
pubmed:dateRevised
2006-11-15
pubmed:meshHeading
pubmed:year
1994
pubmed:articleTitle
The desflurane (Tec 6) vaporizer: design, design considerations and performance evaluation.
pubmed:affiliation
Department of Anesthesia, University of California, San Francisco 94143-0648.
pubmed:publicationType
Journal Article, Research Support, Non-U.S. Gov't