pubmed:abstractText
The Tono-Pen XL (Bio-Rad, Santa Ana, CA) was compared with manometer readings for intraoperative measurement of intraocular pressure (IOP) in 40 eyes of 40 consecutive patients after vitrectomy, lensectomy, and fluid-air exchange. Tono-Pen readings corresponding to manometer readings of 10, 20, 30, 40, and 50 mmHg were obtained in a masked fashion with a randomized sequence of manometer readings. A correlation was obtained between the manometer and Tono-Pen readings (r = 0.96 in emmetropic eyes and r = 0.93 in myopic eyes). The regression curve that represents the calibration curve of Tono-Pen in terms of the manometer readings for air-filled vitrectomized eyes was obtained. Any Tono-Pen reading can be easily translated into the corresponding manometer reading by referring to the curve. The Tono-Pen can therefore be effectively used to accurately determine intraoperative IOP in eyes undergoing vitrectomy, lensectomy, and fluid-air exchange.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:status
MEDLINE
pubmed:issn
0275-004X
pubmed:author
pubmed:issnType
Print
pubmed:volume
13
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
307-11
pubmed:dateRevised
2006-12-15
pubmed:meshHeading
pubmed:year
1993
pubmed:articleTitle
Intraoperative measurement of intraocular pressure in vitrectomized aphakic air-filled eyes using the Tono-Pen XL.
pubmed:affiliation
Vitreoretinal Department, Sankara Nethralaya, Medical Research Foundation, Madras, India.
pubmed:publicationType
Journal Article, Comparative Study, Research Support, Non-U.S. Gov't