pubmed:abstractText |
We evaluated the analytical and clinical performance of an immunoassay for cardiac troponin T (cTnT). Within-run and total imprecision ranged from 1.6 to 11.3%. The sensitivity and linear range was 0.015 and 13 micrograms/L, respectively. Frozen samples were stable for at least 8 weeks. No interferences were seen with lipids or bilirubin (total and conjugated). Hemoglobin caused a negative bias at concentrations > 4 g/L. Heparinized plasma showed a 6% negative bias compared with serum. The clinical utility of cTnT was compared with that of creatine kinase (CK)-MB (mass assay). The sensitivity of cTnT measurements from 63 patients with acute myocardial infarction (AMI) (cTnT cutoff 0.1 microgram/L) was 60% at 0-3 h, 59% at 3-6 h, 94% at 6-9 h, 90% at 9-12 h, 99% at 12-24 h, 92% at 24-48 h, 83% at 48-72 h, and 100% at 72-96 h. Corresponding results for CK-MB (cutoff 5.0 micrograms/L and 2.5% relative index) were 45%, 64%, 82%, 97%, 87%, 81%, 54%, and 59%. The specificity of the markers from 49 non-AMI patients was 46% and 79% for cTnT and CK-MB, respectively. We show that CK-MB is more specific for diagnosis of AMI, and propose that cTnT is more sensitive to myocardial injury.
|
pubmed:commentsCorrections | |
pubmed:language |
eng
|
pubmed:journal | |
pubmed:citationSubset |
IM
|
pubmed:chemical |
http://linkedlifedata.com/resource/pubmed/chemical/Bilirubin,
http://linkedlifedata.com/resource/pubmed/chemical/Creatine Kinase,
http://linkedlifedata.com/resource/pubmed/chemical/Hemoglobins,
http://linkedlifedata.com/resource/pubmed/chemical/Heparin,
http://linkedlifedata.com/resource/pubmed/chemical/Isoenzymes,
http://linkedlifedata.com/resource/pubmed/chemical/Lipids,
http://linkedlifedata.com/resource/pubmed/chemical/Troponin,
http://linkedlifedata.com/resource/pubmed/chemical/Troponin T
|
pubmed:status |
MEDLINE
|
pubmed:month |
Jun
|
pubmed:issn |
0009-9147
|
pubmed:author | |
pubmed:issnType |
Print
|
pubmed:volume |
40
|
pubmed:owner |
NLM
|
pubmed:authorsComplete |
Y
|
pubmed:pagination |
900-7
|
pubmed:dateRevised |
2006-11-15
|
pubmed:meshHeading |
pubmed-meshheading:8087984-Bilirubin,
pubmed-meshheading:8087984-Creatine Kinase,
pubmed-meshheading:8087984-Drug Stability,
pubmed-meshheading:8087984-False Negative Reactions,
pubmed-meshheading:8087984-Freezing,
pubmed-meshheading:8087984-Hemoglobins,
pubmed-meshheading:8087984-Heparin,
pubmed-meshheading:8087984-Humans,
pubmed-meshheading:8087984-Immunoassay,
pubmed-meshheading:8087984-Isoenzymes,
pubmed-meshheading:8087984-Lipids,
pubmed-meshheading:8087984-Myocardial Infarction,
pubmed-meshheading:8087984-Myocardium,
pubmed-meshheading:8087984-Reference Values,
pubmed-meshheading:8087984-Sensitivity and Specificity,
pubmed-meshheading:8087984-Time Factors,
pubmed-meshheading:8087984-Troponin,
pubmed-meshheading:8087984-Troponin T
|
pubmed:year |
1994
|
pubmed:articleTitle |
Cardiac troponin-T immunoassay for diagnosis of acute myocardial infarction.
|
pubmed:affiliation |
Department of Pathology, Hartford Hospital, CT.
|
pubmed:publicationType |
Journal Article,
Comparative Study,
Research Support, Non-U.S. Gov't
|