pubmed:abstractText
To expand the usage of the centrifugal pump (CP) in open heart surgery, we performed two studies. In the first, we evaluated pulsatile flow in the CP. In vitro pump performance of the Terumo Capiox pump (TCP) and the Sarns Delphin pump (SDP) and increase of free hemoglobin (mg/dl) after driving 6 h were investigated using bovine blood. A roller pump (RP) was used as a comparison. Equally effective pulsatile flow was obtained in both CPs. Hemolysis was less severe in TCP (120 mg/dl) than SDP (210 mg/dl) and RP (320 mg/dl). In the second study, we evaluated a simple flow control method. Flow rate was easily controlled with step-wise clamping of 3-pronged tubing (Triple-flow) without changing rotational speed, regardless of afterload. Fluctuation of flow was much less with this method than with the rotational speed change method. The use of pulsatile flow of TCP, with its minimum increase of hemolysis and the easier flow control method during the weaning process, may expand the usage of CP in open heart surgery.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:chemical
pubmed:status
MEDLINE
pubmed:month
Sep
pubmed:issn
0160-564X
pubmed:author
pubmed:issnType
Print
pubmed:volume
18
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
643-9
pubmed:dateRevised
2006-11-15
pubmed:meshHeading
pubmed:year
1994
pubmed:articleTitle
Pulsatile flow and simple flow control method during weaning period in centrifugal pump: toward more expanded usage in open heart surgery.
pubmed:affiliation
Department of Cardiovascular Surgery, Heart Institute of Japan, Tokyo Women's Medical College.
pubmed:publicationType
Journal Article, Comparative Study