A technique is described to modify a standard fluorescence microscope for time-resolved visualization of delayed luminescing substances with decay times from 50 microseconds to several milliseconds. The modification consists of synchronized operation of a mechanical shutter, positioned in an aperture plane in the excitation pathway, simultaneously with a ferro-electric liquid crystal (FLC) shutter on the emission side. Operation of the microscope is through a microprocessor interfaced keypad by which all timing parameters can be adjusted for optimal suppression of fast decaying luminescence. Accuracy of the timing was within 1 microsecond. Prompt fluorescence was suppressed up to 10(6) times, as determined for bright prompt fluorescing microspheres. The use of the FLC shutter resulted in a reduction in emission intensity by a factor of 8 (due to the use of polarizers, the lower transmission of the FLC devices, and IR blocking filters). No significant image degradation due to shutter operations was observed. The modified microscope was successfully used for the visualization of delayed luminescing immunolabels, such as inorganic phosphor particles and lanthanide chelates, as well as naturally phosphorescing materials.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:status
MEDLINE
pubmed:month
Jun
pubmed:issn
0196-4763
pubmed:author
pubmed:issnType
Print
pubmed:day
1
pubmed:volume
16
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
113-7
pubmed:dateRevised
2006-11-15
pubmed:meshHeading
pubmed:year
1994
pubmed:articleTitle
Use of ferro-electric liquid crystal shutters for time-resolved fluorescence microscopy.
pubmed:affiliation
Department of Cytochemistry and Cytometry, University of Leiden, The Netherlands.
pubmed:publicationType
Journal Article, Research Support, Non-U.S. Gov't