pubmed:abstractText
Evaluation has been performed of compensators generated by means of a computerized three-dimensional treatment planning system that can utilize either digitized slice profiles or CT scans. Two methods of calculating compensator thickness are used: the modified Batho power law (dSAR) method for digitized profiles and the equivalent TAR (eqTAR) method for CT scans. This system not only compensates for patient surface contours but also compensates for internal inhomogeneities. In addition, any required wedging will be incorporated in the compensator generation. This system has been tested for a number of extreme cases with inhomogeneities and sloping contours. Good agreement was obtained between the measured and computer calculated dose profiles especially along the central axis of the beam. A "Profile Uniformity Index" was defined to quantify the goodness of dose compensation in three dimensions. Compensation using this system can achieve good dose uniformity within the target volume in all clinical cases and is definitely an improvement over systems based solely on tissue deficit.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:status
MEDLINE
pubmed:issn
0958-3947
pubmed:author
pubmed:issnType
Print
pubmed:volume
19
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
227-35
pubmed:dateRevised
2007-11-15
pubmed:meshHeading
pubmed:year
1994
pubmed:articleTitle
Description and evaluation of a new 3-D computerized treatment planning dose compensator system.
pubmed:affiliation
Ontario Cancer Institute/Princess Margaret Hospital, Department of Clinical Physics, Toronto, Canada.
pubmed:publicationType
Journal Article