pubmed-article:7850235 | pubmed:abstractText | The characteristics of fluorescent essay of total and effective albumin concentrations by Zond-Albumin kit (experimental lot) were as follows: intraseries reproducibility in measurements not more than 1.6%, the error in serum dosage being mainly responsible for the total error. After unsealing the ampules the reagents may be stored at 5 degrees C for at least a week. Fluorescence intensity recorded in measurements in the majority of samples is stable for at least 50 min after addition of reagents. Measured values of total albumin concentration coincide with the results of measuring mass albumin concentration by bromocresol blue method, coefficient of linear correlation r = 0.98. | lld:pubmed |