Statements in which the resource exists as a subject.
PredicateObject
rdf:type
lifeskim:mentions
pubmed:issue
11 Pt 2
pubmed:dateCreated
1995-3-6
pubmed:abstractText
Minute volume is a truly physiological sensor for rate adaptive pacing that correlates with metabolic expenditure throughout the range of physical activity. Criticism has centered on the slow initial response compared to less physiological sensors. A new algorithm, consisting of rate augmentation factor and programmable speed of response, has been incorporated in the 1206 META III pacemaker generator and was designed to improve the rate response at lower levels of exertions. Rate augmentation factor increases the programmed rate response factor by 3, 6, or 10 when set to low, medium, or high, respectively; this augmentation lasting to 50% of the maximum programmed rate. Response time can be programmed to medium or fast.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:status
MEDLINE
pubmed:month
Nov
pubmed:issn
0147-8389
pubmed:author
pubmed:issnType
Print
pubmed:volume
17
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
1960-5
pubmed:dateRevised
2004-11-17
pubmed:meshHeading
pubmed:year
1994
pubmed:articleTitle
New algorithms to increase the initial rate response in a minute volume rate adaptive pacemaker.
pubmed:affiliation
Department of Cardiological Sciences, St. George's Hospital Medical School, London, United Kingdom.
pubmed:publicationType
Journal Article, Clinical Trial, Randomized Controlled Trial