pubmed:abstractText
The Biotrack 516 is a simple, automated whole blood phenytoin (PHT) assay that reports corresponding total serum concentrations in 3 min. We compared Biotrack results in 58 patients with the total and unbound serum PHT concentrations measured by the standard TDx fluorescence polarization immunoassay. Correlation with total TDx concentration was high (r = 0.98); median absolute error was 1.4 micrograms/ml. Correlation of unbound PHT with Biotrack (r = 0.95) was comparable to correlation of unbound and total TDx (r = 0.94). The Biotrack assay is a promising method for clinical monitoring of PHT concentrations.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:chemical
pubmed:status
MEDLINE
pubmed:month
Nov
pubmed:issn
0013-9580
pubmed:author
pubmed:issnType
Print
pubmed:volume
36
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
1152-4
pubmed:dateRevised
2007-11-15
pubmed:meshHeading
pubmed:year
1995
pubmed:articleTitle
Near-patient rapid assay of phenytoin concentration.
pubmed:affiliation
Department of Neurology, Yale University School of Medicine, New Haven, Connecticut, USA.
pubmed:publicationType