We tested the localization accuracy of magnetoencephalography (MEG) and electrocorticography (ECoG) for a current dipole in a saline filled sphere at depths ranging from 1 to 6 cm at 1 cm intervals. We used standard neuromagnetometer placements and subdural electrode grids, previously employed for patient studies, with precise measurements of sensor and electrode locations with a 3-dimensional spatial digitizer. MEG and ECoG had comparable accuracy with mean errors of 1.5 and 1.8 mm, respectively. It appears that use of the spatial digitizer increases accuracy for both MEG and EGoG localizations. The larger errors in the ECoG with increasing depths could be attributed to under-sampling of the spatial pattern of the field which spreads out with deeper sources. It should be noted that in clinical applications a grid of the dimensions used here would most typically be used for superficial sources on the cortex with depth recordings being preferred for investigations of deep epileptogenic activity. Results are encouraging for continued development of non-invasive MEG methods for further definition of epileptogenic zones in the brain.
pubmed:grant
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:status
MEDLINE
pubmed:month
Feb
pubmed:issn
0013-4694
pubmed:author
pubmed:issnType
Print
pubmed:volume
94
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
109-14
pubmed:dateRevised
2008-9-9
pubmed:meshHeading
pubmed:year
1995
pubmed:articleTitle
MEG and ECoG localization accuracy test.
pubmed:affiliation
Department of Electrical Engineering, University of California, Los Angeles.