Statements in which the resource exists as a subject.
PredicateObject
rdf:type
lifeskim:mentions
pubmed:issue
1
pubmed:dateCreated
1981-2-26
pubmed:abstractText
The effectiveness and safety of the defibrillating effect of a bipolar impulse produced by a Soviet DK1-01 defibrillator and the Monopolar impulse Ed-Mark have been tested in 14 experiments on the anaesthetized dogs. A quantitative criterion to compare the safety of different defibrillators and an electro-therapeutic index of the defibrillating action are being suggested. It is defined as the ratio of the thresholds of energy in joules causing lesions to that causing defibrillation. In the above experiments the mean value of this therapeutic index was 1.59 +/- 1.41 for the Ed-Mark impulse and 4.32 +/- 2.78 for the defibrillator DK1-01 impulse. The difference of the mean values was statistically significant (P = 0.02).
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:status
MEDLINE
pubmed:month
Mar
pubmed:issn
0300-9572
pubmed:author
pubmed:issnType
Print
pubmed:volume
8
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
53-67
pubmed:dateRevised
2009-8-25
pubmed:meshHeading
pubmed:year
1980
pubmed:articleTitle
Criteria of efficiency and safety of the defibrillating impulse.
pubmed:publicationType