pubmed:abstractText |
A surgery of 349 cardiac surgeons showed that during a six-year period, a pump oxygenator accident serious enough to cause patient injury or death occurred one per 1,000 procedures. A total of 264 deaths occurred as a direct results of an accident. Air embolism and disseminated intravascular coagulation were the two most common problems. Low-level alarm systems were reported to be used by 42% of the respondents and activated clotting times were used by 63%. Rigorous use of alarm systems and heparin monitoring could reduce the incidence of pump-related accidents.
|
pubmed:language |
eng
|
pubmed:journal | |
pubmed:citationSubset |
AIM
|
pubmed:chemical | |
pubmed:status |
MEDLINE
|
pubmed:month |
Apr
|
pubmed:issn |
0003-4975
|
pubmed:author | |
pubmed:issnType |
Print
|
pubmed:volume |
29
|
pubmed:owner |
NLM
|
pubmed:authorsComplete |
Y
|
pubmed:pagination |
336-40
|
pubmed:dateRevised |
2004-11-17
|
pubmed:meshHeading |
pubmed-meshheading:7362326-Aged,
pubmed-meshheading:7362326-Blood Volume,
pubmed-meshheading:7362326-Cardiopulmonary Bypass,
pubmed-meshheading:7362326-Embolism, Air,
pubmed-meshheading:7362326-Female,
pubmed-meshheading:7362326-Heart Diseases,
pubmed-meshheading:7362326-Heparin,
pubmed-meshheading:7362326-Humans,
pubmed-meshheading:7362326-Male,
pubmed-meshheading:7362326-Middle Aged,
pubmed-meshheading:7362326-Oxygenators,
pubmed-meshheading:7362326-Partial Thromboplastin Time
|
pubmed:year |
1980
|
pubmed:articleTitle |
Air embolism and other accidents using pump oxygenators.
|