pubmed:abstractText |
Anderson Technical, Inc. has developed rapid, semiautomated equipment for antimicrobial susceptibility testing. A total of 310 fresh clinical bacterial isolates (226 gram-negative and 75 gram-positive) were tested with the Anderson Technical system and compared with those of Micro-Media Systems, Inc. For the gram-positive organisms, 74.3% of the test pairs had identical minimum inhibitory concentration values, whereas 99.1 and 0.9% of the test pairs had minimal inhibitory concentration values differing by less than or equal to 1 and greater than 1 dilution level, respectively. Identical minimum inhibitory concentration values were obtained for 67.2% of the gram-negative test pairs, whereas 97.6 and 2.4% differed by less than or equal to 1 and greater than 1 dilution level, respectively. For all organisms tested, 98.0% differed by less than or equal to 1 dilution level. The Anderson Technical equipment proved to be a rapid and flexible system for microdilution testing.
|
pubmed:commentsCorrections |
http://linkedlifedata.com/resource/pubmed/commentcorrection/7325643-1147582,
http://linkedlifedata.com/resource/pubmed/commentcorrection/7325643-457852,
http://linkedlifedata.com/resource/pubmed/commentcorrection/7325643-5504607,
http://linkedlifedata.com/resource/pubmed/commentcorrection/7325643-626491,
http://linkedlifedata.com/resource/pubmed/commentcorrection/7325643-7372803,
http://linkedlifedata.com/resource/pubmed/commentcorrection/7325643-7381010,
http://linkedlifedata.com/resource/pubmed/commentcorrection/7325643-7419705
|
pubmed:language |
eng
|
pubmed:journal | |
pubmed:citationSubset |
IM
|
pubmed:chemical | |
pubmed:status |
MEDLINE
|
pubmed:month |
Dec
|
pubmed:issn |
0066-4804
|
pubmed:author | |
pubmed:issnType |
Print
|
pubmed:volume |
20
|
pubmed:owner |
NLM
|
pubmed:authorsComplete |
Y
|
pubmed:pagination |
777-82
|
pubmed:dateRevised |
2009-11-18
|
pubmed:meshHeading | |
pubmed:year |
1981
|
pubmed:articleTitle |
Laboratory evaluation of the Anderson Technical semiautomated susceptibility testing system.
|
pubmed:publicationType |
Journal Article
|