Several authors have suggested using heavy metal filters with K edges in the diagnostic energy range to reduce the width of the x-ray spectrum and hence reduce patient radiation exposure. This spectral narrowing also increases subject contrast and permits an increase in tube potential. Results of contrast measurements are presented for a 250 mu gadolinium filter. It was found that aluminum filter contrast could be matched by using 8 to 10 kVp higher potential with the gadolinium filter. Similar results were found for calcium tungstate and rare-earth screens. Measurements were also done to determine skin exposure and mAs ratios for both constant contrast and constant kVp technique conversion methods. A simple theory with one adjustable parameter gives a reasonable fit to the experimental results.
|