pubmed:abstractText
A high-performance liquid chromatographic method for the measurement of nifuroxazide in plasma is described. The technique is based on the single extraction of the drug from buffered plasma with chloroform, using nifuratel as internal standard. The chromatographic system consisted of a 15 cm x 4.6 mm I.D. stainless-steel column packed with Spherisorb ODS, 5 micrometer, and the mobile phase was acetonitrile-orthophosphoric acid (pH 2.5) (30:70). The method was able to measure accurately plasma nifuroxazide concentrations down to 2 ng . ml-1 using 2 ml of sample with no interference from endogenous compounds. The coefficients of variation of the method at 200 and 2 ng . ml-1 were 3% and 15%, respectively, and the calibration graph was linear in this range. The use of automatic injection makes the method suitable for the routine analysis of large numbers of samples.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:chemical
pubmed:status
MEDLINE
pubmed:month
Jan
pubmed:issn
0021-9673
pubmed:author
pubmed:issnType
Print
pubmed:day
16
pubmed:volume
204
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
329-33
pubmed:dateRevised
2004-11-17
pubmed:meshHeading
pubmed:year
1981
pubmed:articleTitle
Determination of nifuroxazide in biological fluids by automated high-performance liquid chromatography with large-volume injection.