pubmed:abstractText
Hypo-test (Hypoguard Ltd., Suffolk, United Kingdom), a new semiquantitative audio urine-glucose analyzer for blind diabetic individuals, was tested for accuracy and reproducibility against a laboratory glucose method, and for ease of operation by two blind diabetic subjects. Hypo-test results were accurate for urine glucose in the range 0-0.75%, but there was a pronounced tendency for results to be too low at higher urine-glucose concentrations. The reproducibility at high urine-glucose concentrations was poor, and the meter was too sensitive to delays or decreases in the prescribed timing procedure. However, the two blind diabetic subjects found the instrument easy to operate.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:status
MEDLINE
pubmed:issn
0149-5992
pubmed:author
pubmed:issnType
Print
pubmed:volume
5
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
430-2
pubmed:dateRevised
2004-11-17
pubmed:meshHeading
pubmed:articleTitle
Evaluation of Hypo-test, a semiquantitative audio urine-glucose analyzer for blind diabetic individuals.