Statements in which the resource exists as a subject.
PredicateObject
rdf:type
lifeskim:mentions
pubmed:issue
6
pubmed:dateCreated
1982-8-7
pubmed:abstractText
We evaluated an incentive spirometer (IS) for monitoring changes in lung function in hospitalized patients. Accuracy and reproducibility of IS measurements of known volumes were adequate (r = 0.87). Flow dependency was demonstrated but was not significant in the clinically useful range. Reproducibility of IS measurements in five normal subjects was good, with a small training effect uncovered. In 15 patients with asthma and chronic obstructive lung disease, change in IS values closely correlated with spirometrically measured changes in volume and flows (best correlation: IS versus FEV1/FVC%, r = 0.98) and in peak flow. The performance of the IS as tested and its availability in most hospitals outweigh its limitations. We advocate its use as an adjunct in monitoring progress of hospitalized patients with obstructive lung disease.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
AIM
pubmed:status
MEDLINE
pubmed:month
Jun
pubmed:issn
0038-4348
pubmed:author
pubmed:issnType
Print
pubmed:volume
75
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
687-90, 693
pubmed:dateRevised
2004-11-17
pubmed:meshHeading
pubmed:year
1982
pubmed:articleTitle
Incentive spirometer for bedside studies.
pubmed:publicationType
Journal Article