pubmed:abstractText
We evaluated the performance of the Radiometer TCM-1 Transcutaneous Oxygen Monitor in a Class III neonatal-intensive-care unit during an 11-month clinical trial (in excess of 4000 h) in which we monitored 115 infants ranging in gestational age from 25 to 43 weeks and in birthweight from 595 to 4220 g. In vitro transcutaneous partial pressure of oxygen (ptcO2) measurements were accurate within 1% of the theoretical value and precise (1.4% CV at 156 mmHg) at 44 degrees C. Determinations at other temperature were less accurate except in rare instances when specific patient-associated ptcO2/paO2 bias was identified. In vitro comparison of ptcO2 (y) and arterial partial pressure of oxygen (paO2) (x) measurements agreed well (y = 0.90x + 6.9, r = 0.94) except when infants were less than 24 h old. In this latter group the ptcO2/paO2 ratio usually exceeded 1.0 and fluctuated unpredictably. ptcO2 measurements on infants less than 24 h old had to be interpreted cautiously. In general, the ptcO2 and paO2 measurements agreed sufficiently well for the physicians to use them as the basis for directing patient care.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:chemical
pubmed:status
MEDLINE
pubmed:month
Apr
pubmed:issn
0009-9147
pubmed:author
pubmed:issnType
Print
pubmed:volume
26
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
629-33
pubmed:dateRevised
2007-11-15
pubmed:meshHeading
pubmed:year
1980
pubmed:articleTitle
Performance of a radiometer transcutaneous oxygen monitor in a neonatal-intensive-care unit.
pubmed:publicationType
Journal Article, Clinical Trial