pubmed:abstractText
We have developed an in vitro haemoperfusion circuit, which closely simulates the conditions encountered in a clinical haemoperfusion, for testing the blood compatibility of new adsorbent materials. Fresh heparinised human blood is perfused for 2 hours over two model columns arranged in parallel closed circuits. The present study was performed to investigate the blood compatibility of human serum albumin-coated Amberlite XAD-7 resin. Mean platelet losses over 2 hours were greater following perfusion of the uncoated resin (44 +/- SE 5.6%) than the albumin-coated resin (17 +/- SE 2.2%) (p less than 0.01). Mean white cell losses were similar (55%) for both resins. No rises in Swank screen filtration pressure, which detects the presence of cellular aggregates in blood, were observed. Coating of XAD-7 resin with HSA therefore improves its blood compatibility with respect to platelet losses and we thus intend to use this preparation in a clinical trial of resin haemoperfusion in fulminant hepatic failure.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:chemical
pubmed:status
MEDLINE
pubmed:month
May
pubmed:issn
0391-3988
pubmed:author
pubmed:issnType
Print
pubmed:volume
1
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
129-34
pubmed:dateRevised
2007-11-15
pubmed:meshHeading
pubmed:year
1978
pubmed:articleTitle
The use of an in vitro haemoperfusion circuit to evaluate the blood compatibility of albumin-coated Amberlite XAD-7 resin.
pubmed:publicationType
Journal Article, Comparative Study, In Vitro