pubmed:abstractText
The methodology employed to calculate radiographic signal-to-noise ratio (SNR) for the commonly used disk-shaped test object is reviewed. Although the "matched" aperture yields the maximum SNR, its shape is difficult to achieve experimentally. On the other hand, a circular aperture having the same size as the object is much simpler to realize. We have performed a numerical analysis of the SNRs obtained with matched and circular apertures for two screen/film systems (a mammographic and a general purpose combination) for a range of test-object sizes. We find that for object sizes greater than 0.5 mm in diameter, there is a less than 4% difference between the SNRs obtained with the two apertures. The shapes of some of the matched apertures are also presented. We conclude that the SNR determined with the circular aperture is a simpler and more useful determinant of system performance for the screen/film combinations studied.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:status
MEDLINE
pubmed:issn
0094-2405
pubmed:author
pubmed:issnType
Print
pubmed:volume
10
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
467-9
pubmed:dateRevised
2000-12-18
pubmed:meshHeading
pubmed:articleTitle
Signal-to-noise ratio considerations in radiographic imaging.
pubmed:publicationType
Journal Article