Statements in which the resource exists as a subject.
PredicateObject
rdf:type
lifeskim:mentions
pubmed:issue
3
pubmed:dateCreated
1982-9-17
pubmed:abstractText
Reusable, corrugated, expiratory limb ventilator tubing that had been in use for 24 hours, were randomly allocated to one of three groups: no treatment (N = 36); detergent wash (N = 83); or a detergent wash followed by a 10 minute immersion in a 1:16 dilution of synergized glutaraldehyde-phenate solution which was reused for 30 days. (Between 10 and 22 tubes were tested in each five day interval during this 30-day period.) Tubes were quantitatively and qualitatively cultured. There were significant differences in both the percent of contaminated tubes (no treatment = 92%, detergent wash = 72%, glutaraldehyde-phenate = 0 to 20%) and numbers of microorganisms per tube (no treatment = 3.2 x 10(6), detergent wash = 1.3 x 10(4), glutaraldehyde-phenate = 0 to 182) between groups. There was no apparent decrease in glutaraldehyde-phenate's efficacy throughout the 30-day reuse period, and in the final five days of the reuse period it was completely effective.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:chemical
pubmed:status
MEDLINE
pubmed:issn
0195-9417
pubmed:author
pubmed:issnType
Print
pubmed:volume
3
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
240-4
pubmed:dateRevised
2010-4-12
pubmed:meshHeading
pubmed:articleTitle
An efficacy evaluation of a synergized glutaraldehyde-phenate solution in disinfecting respiratory therapy equipment contaminated during patient use.
pubmed:publicationType
Journal Article, Clinical Trial, Comparative Study, Randomized Controlled Trial