Statements in which the resource exists as a subject.
PredicateObject
rdf:type
lifeskim:mentions
pubmed:issue
4
pubmed:dateCreated
1982-1-20
pubmed:abstractText
Combined transcutaneous PO2-PCO2 electrodes are described in which the interaction between the two electrodes due to OH- production at the O2 cathode has been eliminated. An anode of either anodized aluminum or platinum has been driven at a current equal to cathode current to force stoichiometric consumption of OH- at its rate of production. The AgCl reference electrode operates at zero current. O2 sensitivity was not significantly altered by electrolyte pH variation from 6.7 to 9.0 with variations by PCO2. These electrodes have been found stable both with and without spacers, and with electrolytes dissolved in 50-100% ethylene glycol. In 22 anesthetized patients, with electrode temperature of 43 degrees C (s refers to skin surface, a to arterial blood); PsO2 = 0.52PaO2 + 15 (range 54-300) (r = 0.66; Sy . x = 29.6; n = 46); and PsCO2 = 1.39PaCO2 + 2.1 (range 24-98) (r = 0.99; Sy . x = 2.28; n = 48).
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:chemical
pubmed:status
MEDLINE
pubmed:month
Oct
pubmed:issn
0161-7567
pubmed:author
pubmed:issnType
Print
pubmed:volume
51
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
1027-32
pubmed:dateRevised
2000-12-18
pubmed:meshHeading
pubmed:year
1981
pubmed:articleTitle
A combined transcutaneous PO2-PCO2 electrode with electrochemical HCO3- stabilization.
pubmed:publicationType
Journal Article