Statements in which the resource exists as a subject.
PredicateObject
rdf:type
lifeskim:mentions
pubmed:issue
2
pubmed:dateCreated
1982-1-9
pubmed:abstractText
We describe a simple direct extraction method for the gas-liquid chromatography determination of serum valproic acid. The working range for the assay is 2-180 mg/L and our within-run precision was 5.8 and 4.3% at the 40 and 90 mg/L concentrations respectively. Hemolyzed and lipemic sera as well as samples from patients with hyperbilirubinemia and from patients with decreased renal function were put through the assay and no interfering peaks were noted. Interference occurred when teflon-lined screw caps were used during the extraction step. The method was proven to be accurate by linear regression analysis of samples containing weighed-in amounts of valproic acid. The above assay was compared to an enzyme immunoassay technique (EMIT). The working range for the latter is 10-150 mg/L and the with-run precision was 10.8 and 5.9% and 90 mg/L concentration respectively. Samples were run by both the gas-liquid chromatograph and enzyme immunoassay methods and gave very similar results over the range 16-139 mg/L.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:chemical
pubmed:status
MEDLINE
pubmed:month
Apr
pubmed:issn
0009-9120
pubmed:author
pubmed:issnType
Print
pubmed:volume
14
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
87-90
pubmed:dateRevised
2006-11-15
pubmed:meshHeading
pubmed:year
1981
pubmed:articleTitle
Comparison of gas-liquid chromatography and EMIT assay for serum valproic acid.
pubmed:publicationType