We compared four instruments, [Sequential Multiple Analyzer Computer (SMAC), AutoAnalyzer II (AAII), StatLyte and PVA-4] for the determination of Na, K, Cl and CO2 using the National Committee for Clinical Laboratory Standard's (NCCLS) protocol (PSEP-1) for evaluation of precision and accuracy of automated instruments. The protocol was applied concurrently to all four instruments. The repeatability results showed SMAC to have the highest precision among-days for K, Cl, and CO2 whereas StatLyte showed the highest precision among-days for Na. The PVA-4 had the best within-run precision for Na and Cl while the AAII and SMAC had the best within-run precision for K and CO2 respectively. Systematic biases among the instruments were small for Na and K. The AAII exhibited a nearly constant bias at all concentrations of Cl relative to the other three instruments. For CO2 the SMAC showed a large proportional bias compared with the other three instruments.
|
pubmed:language |
eng
|
pubmed:journal | |
pubmed:citationSubset |
IM
|
pubmed:chemical |
http://linkedlifedata.com/resource/pubmed/chemical/Bicarbonates,
http://linkedlifedata.com/resource/pubmed/chemical/Carbon Dioxide,
http://linkedlifedata.com/resource/pubmed/chemical/Chlorides,
http://linkedlifedata.com/resource/pubmed/chemical/Electrolytes,
http://linkedlifedata.com/resource/pubmed/chemical/Potassium,
http://linkedlifedata.com/resource/pubmed/chemical/Sodium
|
pubmed:status |
MEDLINE
|
pubmed:month |
Aug
|
pubmed:issn |
0009-9120
|
pubmed:author | |
pubmed:issnType |
Print
|
pubmed:volume |
13
|
pubmed:owner |
NLM
|
pubmed:authorsComplete |
Y
|
pubmed:pagination |
164-70
|
pubmed:dateRevised |
2006-11-15
|
pubmed:meshHeading |
pubmed-meshheading:6778632-Autoanalysis,
pubmed-meshheading:6778632-Bicarbonates,
pubmed-meshheading:6778632-Carbon Dioxide,
pubmed-meshheading:6778632-Chemistry, Clinical,
pubmed-meshheading:6778632-Chlorides,
pubmed-meshheading:6778632-Electrolytes,
pubmed-meshheading:6778632-Humans,
pubmed-meshheading:6778632-Potassium,
pubmed-meshheading:6778632-Quality Control,
pubmed-meshheading:6778632-Sodium
|
pubmed:year |
1980
|
pubmed:articleTitle |
Comparison of four electrolyte instruments using the proposed NCCLS protocol for evaluation of precision and accuracy of automated instruments.
|
pubmed:publicationType |
Journal Article,
Comparative Study
|