A device for the continuous measurement of left ventricular (LV) function was tested in a series of 34 subjects. The instrument consisted of 2 arrays of radiation sensitive cadmium telluride detectors held in place over the region of the left ventricle and lung by a vest-like garment (hence the name VEST). The VEST electronic instrumentation included analog-to-digital converters, a battery pack, microprocessor and gating device, which were worn in a back pack. Data generated by the VEST, including the digitized average electrocardiogram, RR interval, counts/13 ms in each radiation detector, and time since commencement of data recording, were recorded on a cassette tape recorder every 2 minutes for subsequent analysis. At the conclusion of conventional multigated blood pool imaging, the VEST was positioned and worn by the subjects while supine, standing in place and walking. The correlation of ejection fraction calculated independently from the VEST and scintillation camera data was greater than 0.95. The inter-record reproducibility of the ejection fraction measured by the VEST in sedentary subjects was less than 3%.
pubmed:grant
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
AIM
pubmed:status
MEDLINE
pubmed:month
Sep
pubmed:issn
0002-9149
pubmed:author
pubmed:issnType
Print
pubmed:day
1
pubmed:volume
52
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
601-6
pubmed:dateRevised
2007-11-14
pubmed:meshHeading
pubmed:year
1983
pubmed:articleTitle
An ambulatory ventricular function monitor: validation and preliminary clinical results.