pubmed:abstractText
In 6 anesthetized and paralyzed supine dogs, ventilation by high-frequency chest wall compression (HFCWC) was accomplished by a piston pump rapidly oscillating the pressure in a modified double blood pressure cuff wrapped around the lower thorax. Testing applied frequencies at 3, 5, 8, and 11 Hz, applied peak cuff pressures ranged from 30 to 230 cmH2O. This produced swings of esophageal pressure as high as 18 cmH2O and peak oscillatory air flow ranging from 0.7 to 1.6 L/s. Oscillatory tidal volume declined with increasing frequency and ranged from a mean of 61 to 45 ml. After 30 min of applied HFCWC, arterial blood gas determinations revealed a mean PaCO2 of 29.3 mmHg at 5 Hz, 35 mmHg at 3 Hz, 36 mmHg at 8 Hz, and 51 mmHg at 11 Hz. Mean PaO2 improved from ventilator control values at 3 Hz, remained unchanged at 5 and 8 Hz, and declined at 11 Hz. In 2 dogs breathing spontaneously, HFCWC applied at 5 and 11 Hz resulted in a reduction in spontaneous minute ventilation, mainly by a reduction in spontaneous tidal volume, whereas arterial blood gas values changed slightly. One dog ceased to breath spontaneously within 5 min of application of HFCWC as the PaCO2 fell below control values. We conclude that in dogs with normal lungs, HFCWC may assist spontaneous ventilation. In paralyzed dogs, HFCWC may be of sufficient magnitude to cause hyperventilation.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
AIM
pubmed:chemical
pubmed:status
MEDLINE
pubmed:month
Jun
pubmed:issn
0003-0805
pubmed:author
pubmed:issnType
Print
pubmed:volume
127
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
709-13
pubmed:dateRevised
2006-11-15
pubmed:meshHeading
pubmed:year
1983
pubmed:articleTitle
Ventilation by high-frequency chest wall compression in dogs with normal lungs.
pubmed:publicationType
Journal Article, Comparative Study, Research Support, Non-U.S. Gov't